Calibration Logs
Calibration logs are the audit trail of every calibration applied to the device's gas and particle sensors. Each row records the calibration type, the reference period used, the resulting offset constant and factor, and an R² correlation preview.
For the calibration workflow itself, see Cloud calibration.

Filters
| Filter | Description |
|---|---|
| Date/time range | Restrict the logs to a specific period. |
| Calibration Type | Baseline, Span, or All — see the filter options below. |
- ·Baseline — only logs from baseline calibrations (sets the zero-current offset).
- ·Span — only logs from span calibrations (sets the sensitivity factor across a range).
- ·All — show both.
Columns
| Column | Description |
|---|---|
| Date and Time | When the calibration was applied. |
| Type | Baseline or span. |
| Offset Constant | Offset constant applied after a baseline calibration. |
| Offset Factor | Offset factor applied after a span calibration. |
| Baseline Time Start / End | Period used to select the minimum reference value during baseline calibration. |
| Span Time Start / End | Period used to select the maximum reference value during span calibration. |
| R² Correlation Result | Preview of the R² assessment — how well the corrected data tracks the reference data if this calibration is applied. |
The R² preview is indicative, not definitive. It's calculated on the corrected values from the selected time window; it doesn't necessarily match the full R² assessment against all historical values in the database.