Maintenance Procedures
Cette page n'est pas encore disponible en français. La version anglaise est affichée.
This section covers the maintenance procedures that may be needed over the lifetime of an Air Pro 2 in service.
In this section
The EnviroSense and particle module procedures are also available as a printable service instruction: Exchanging the Envirosense and Particle Sensor Modules (PDF).
Screw handling
The enclosure and the particle sensor use 3×12 mm Torx T10 stainless screws in PA12 plastic. The plastic thread strips before the screw breaks, so handle them gently.
| Unscrewing | Tightening | |
|---|---|---|
| Torque | 0.6 Nm or less | 0.4–0.6 Nm |
| Speed | 200 rpm or less, or by hand | 200 rpm or less, or by hand |
| Tool | Hand driver or driver with clutch | Hand driver or driver with clutch |
- ·Never use an impact driver.
- ·Keep the T10 bit fully seated with light pressure, so it does not slip and damage the screw head.
- ·If a screw does not loosen at 0.8 Nm, stop. Do not force it, and contact Sensorbee Support.
- ·When refitting, first turn the screw anticlockwise by hand until it drops into the existing thread, then tighten. This prevents cutting a new thread in the plastic.
- ·Do not overtighten.
- ·Replace the screw if it spins without tightening, or if its head is damaged.