Choosing the location
The location of the unit has a direct effect on the quality of the data. Before you mount anything, check all of these points:
- ·Purpose. Choose the location for the purpose of the data, for example urban background, roadside or the boundary of a construction site.
- ·Free air flow. Nothing may block the air flow within about 0.5 m (20 in) of the air inlet.
- ·No local sources. Keep away from chimneys, flues, ventilation and air-conditioning outlets, generators and vehicles with the engine running. They give wrong gas, particulate and temperature readings.
- ·Height. For most applications, mount the unit 3 to 4 m (10 to 13 ft) above the ground.
- ·Shade for the unit. Direct sun on the enclosure makes the unit warmer inside and affects the readings. Put the unit in the shade if you can.
- ·Sun for the solar panel. A solar panel needs a clear view of the sky. Keep it away from buildings, trees and other things that give shade.
- ·No running water. Do not mount the unit directly under gutters, drainpipes or structures that drip.
- ·Mobile coverage. Check the mobile signal before you decide on the location. See Mobile signal strength.
- ·Distance to people. Keep at least 20 cm (8 in) between the unit and any place where a person stays for a long time. See Radio-frequency exposure.
- ·Access. It must be possible to do maintenance safely, without blocking walkways or traffic.
When the location is chosen, continue with Mounting.