Measurement data and MCERTS
This page explains what type of data the Pro2 gives and what you can use it for. It applies to all measurements.
Five things that give you good data. A good location: free air flow, no local pollution sources, shade for the unit (see Choosing the location). Enough power for the whole year (see Power options). A stable mobile signal (see Mobile signal strength). Time for the sensors to become stable after you switch on. Regular checks, calibration and new modules when needed (see Maintenance).
NOTICE. The Pro2 gives indicative measurements. The Pro2 is not a reference-method instrument. It is not a reference or equivalent method under European ambient air quality legislation or its national equivalents. It is not a U.S. EPA Federal Reference Method or Federal Equivalent Method instrument under 40 CFR Part 53. Do not use its data as the only basis to decide regulatory compliance, to assess occupational exposure, or to make decisions that affect health or safety. The accuracy depends on correct location, installation, maintenance and calibration as described in this documentation, and on the conditions in the surrounding air.
Particulate matter and MCERTS
The PM module is an optical particle counter with a heated inlet. It counts particles from 0.3 to 10 µm. It reports the mass concentration of PM1, PM2.5 and PM10 from 0 to 1000 µg/m³.
The Sensorbee Pro2 is certified under MCERTS as an indicative ambient particulate monitor for PM2.5 and PM10 (certificate CSA MC250462/00; see Regulatory compliance). MCERTS is the Monitoring Certification Scheme of the Environment Agency in the United Kingdom. The certificate has these conditions of use:
- ·Indicative, not reference. In the tests of the scheme, the monitor showed a stated uncertainty against a reference method. It may not be used in national automatic monitoring networks that report against the Ambient Air Quality Directives.
- ·Scope. The certificate covers particulate matter with the composition and concentration seen in the field test. This is typical for urban background and traffic locations.
- ·Five-minute data. The certificate applies to five-minute averages that are uploaded every five minutes. This is the factory setting (see Reporting interval).
- ·Qualitative use with factory calibration. For trends, alerts, source identification and dust management, the certificate allows the factory calibration alone. Describe the data as indicative.
- ·On-site calibration for quantitative use. To report results within the ±50 % uncertainty of the standard, calibrate each unit on site for two weeks against a standard or equivalent method. Then apply the slope and intercept from that calibration. Repeat this at least every 12 months, and every time you move the unit to another site.
- ·Certified configuration. The certificate applies to the Pro2 with the MCERTS-compatible PM module (SB4102). Replace the module only with the same type (see Replacing the particulate matter module). Sensorbee recommends a new module every 24 months. Do not change the inlet.
- ·PM1 is reported, but it is not part of the certificate.
If you need a reference-grade value, place a reference instrument next to the Pro2 and correct the Pro2 data against it in the data platform.
Gas modules
- ·Electrochemical gas sensors need time to become stable after you switch on: several hours in most cases, and longer for a new module. Readings before that are not valid.
- ·Gas modules are consumable parts. Their sensors drift and wear out with time. See Calibration and Installing or replacing gas modules.
- ·Range, resolution, detection limit and operating conditions are different for each gas. The catalogue lists them for each module.
Noise, vibration and weather
The noise module and the external sensors report as described in their own datasheets. If your monitoring programme refers to a standard, check that the sensor you choose meets the class or performance that the standard requires.
Data continuity and time stamps
- ·The unit stores about seven days of measurements. After a connection loss, it sends the stored readings in time order. A short loss of coverage therefore leaves no gap in the data. The data arrives late, and the receiving system must accept that.
- ·If the connection is lost for more than about seven days, data is lost.
- ·Measurements follow the UTC clock. With the five-minute setting, the unit measures at :00, :05, :10 and so on. All units in a group therefore report at the same times.