Troubleshooting
| Problem | Possible cause | What to do |
|---|---|---|
| The LED does not light when you switch on | The battery is disconnected. Or it is empty and too little power comes in. | Connect an external power supply. Check the BAT connector (see Replacing the battery). A solar panel needs direct sunlight. |
| The LED is steady red | Error during start or while connecting to the server | Restart the unit. If the error stays, check the pairing, the server details (URL, security mode, pre-shared key) and the registration on the mobile network. |
| The LED blinks red every 10 s | Low battery. The unit is in deep sleep. | Give the unit charging power. Check that the power source is large enough for the site, the season and the reporting interval. |
| The LED stays blue | The unit cannot register on a mobile network | Check the coverage and the RSRP. Check that the SIM is active and may roam on a local operator. Set the APN if the SIM needs one (see Modem menu). |
| The LED keeps blinking purple | The unit is not paired, or it cannot reach the bootstrap server | Pair the unit. With your own server: submit the pairing again if more than one hour has passed. |
| The unit registers and then stops sending data | Wrong details for the data server, or the keys do not match | Check the server URL, security mode and pre-shared key on the platform. Pair the unit again. |
| The unit does not switch off | External power is still connected | Disconnect external power. Then hold the power button for 4 seconds. |
| The LED is steady white and no data arrives | Menu mode: a USB-C cable is connected | Remove the USB-C cable and restart the unit. |
| An external sensor is missing on the platform | The sensor was connected after the unit started. Or a cable or connector is faulty. Or the unit and the sensor use different baud rates. | Restart the unit. Check cables and connectors. Try the sensor without extender or splitter. Check the sensor supply voltage and the baud rate (see System settings). |
| Gas readings look wrong or drift | The sensor is not stable yet, is at the end of its life, or needs calibration | Wait several hours after you switch on. Compare with a reference. Replace the module (see Installing or replacing gas modules). |
| Particulate readings look wrong | The inlet is blocked, the conditions are outside those described in Measurement data, or the module is at the end of its life | Check that the inlet is clear and that the unit is the right way up. Replace the module (see Replacing the particulate matter module). |
| Data is missing for some periods | The network was not available for longer than the data memory lasts | The unit sends stored data after it connects again. Gaps of more than about seven days cannot be recovered. |
| Water inside the enclosure | The enclosure was not closed correctly, a cable was caught in the seal, or a connector had no cap | Take the unit out of service and contact support. |
For the meaning of every LED colour and pattern, see Status LED. For a weak mobile signal, see Mobile signal strength.